R&S®LMG600 series
High-precision power analyzer (DualPath architecture, 0.015 % of reading accuracy, 1/4/7 channels)

  • Category: Power analyzer
  • Manufacturer: Rohde & Schwarz
  • Best power accuracy: 0.015% rdg. + 0.01% rng.
  • Maximum bandwidth: 10 MHz (S/A channels)
  • Maximum input voltage: 1000 V (direct); 1500 V DC (with adapter)
  • Maximum input current: 32 A (direct)
  • Maximum sample rate: 1.21 MS/s
  • ADC resolution (S/A/C channels): 18 bit
  • ADC resolution (B channel): 16 bit
  • Voltage input impedance: 2.69 MΩ ∥ 4 pF
  • Harmonics analysis order: up to the 2000th order (to EN 61000-4-7)
  • Measuring cycle: 10 ms (minimum)
  • Remote interfaces: USB, LAN, RS-232
  • Rated power consumption: 200–400 W (model dependent)
  • Data sheet: Download PDF
  • Product brochure: Download PDF
  • External link: Official product page

Product overview

The R&S®LMG600 series power analyzer uses a distinctive DualPath architecture that runs broadband (unfiltered) and narrowband (anti-aliased) measurements at the same time, with bandwidth up to 10 MHz and a class-leading 0.015 % of reading accuracy. An 18 bit A/D converter, a 10.1" touchscreen and data logging are built in, the S, A, B and C channel types mix freely in one chassis, and several mainframes chain over a synchronization cable into a virtual analysis system of more than 8 channels. The series is at home in demanding work: development verification, EMC compliance (IEC/EN 61000-3-2/-12 harmonics, EN 61000-3-3/-11 flicker, EN 50564/IEC 62301 standby power) and electric drive systems — an all-round measurement platform for the power electronics engineer.

Key features

  • DualPath architecture: the S channel carries two independent A/D converters and measures broadband (unfiltered, DC–10 MHz) and narrowband (anti-aliased, DC–10 MHz) at the same time, removing the blind spots of a conventional sampling architecture
  • Flexible channel mix: the S, A, B and C channel types combine freely in one chassis, covering both high accuracy and high bandwidth
  • Plug 'n' Measure PCT sensor integration: plug in a PCT precision current transformer and it is recognized automatically with its calibration coefficients transferred, no manual setup, for currents from 200 A to 2000 Arms
  • Bi-directional CAN interface: shares up to 128 measured values onto the CAN bus and responds to 128 trigger conditions, for straightforward integration into vehicle and industrial bus systems
  • LMG Test Suite PC software: automated compliance testing to IEC/EN 61000-3-2/-12 (harmonics), EN 61000-3-3/-11 (flicker) and EN 50564/IEC 62301 (standby power)
  • Process signal interface (PSI): 8 digital I/O, 2 high-speed analog inputs (150 kS/s), 8 low-speed analog inputs and 32 analog outputs, fully synchronized with the power channels
  • Multi-unit synchronization: chain several mainframes over a synchronization cable into a virtual analysis system of more than 8 channels, for large multiphase power assessments
  • Custom menu: an application-specific screen with pass/fail indicators, wiring diagrams and operator input fields, so the test operator needs less training
  • Gapless dynamic range: 500 µA–32 A current and 3 mV–1000 V voltage, at 18 bit resolution with a 10 ms minimum measuring cycle

Typical applications

  • Power electronics development verification (efficiency of inverters, chargers and power supplies)
  • EMC compliance testing (IEC/EN 61000-3-2/-12 harmonics, EN 61000-3-3/-11 flicker, EN 50564/IEC 62301 standby power)
  • Electric drive systems (e-axle, motor controllers, motor efficiency maps)
  • Multiphase power system analysis (three-phase inverters, grid simulation, renewable energy)
  • High-reliability power measurement in aerospace and defense
  • Integration into automated test lines (ATE) in manufacturing
Configurator

Configure your R&S®LMG600 Series

Pick a base model and its options, watch your order list build as you go, then send it for a quote in one click. Every order number (Order No.) comes from Rohde & Schwarz factory data.

1

Base model

6 models
1 / 6
2

Hardware options

9 options
2 / 6
3

Software options (K-options)

6 options
3 / 6
4

Probes

4 options
4 / 6
5

PC software

3 options
5 / 6
6

Accessories

8 options
6 / 6

Available models / variants

This series comes in the 6 model variants below. Click Request a quote to ask KanTek for the full specifications and a price for the model you need.

Model codeSpecification differences
LMG611-A3 (1603.5970.02)Single channel, high accuracy and wide bandwidth, for precision R&D measurementsRequest a quote →
LMG611-B2 (1603.0049.02)Single channel, general purpose, for factory production testingRequest a quote →
LMG611-C2 (1603.5905.02)Single channel, high accuracy at low frequency, for mains and low-order harmonics analysisRequest a quote →
LMG641-BAS (1603.0003.02)Four channels, compact, for three-phase power and synchronous multichannel measurementRequest a quote →
LMG671-BAS (1603.5892.02)Seven channels, flagship, for a full efficiency assessment of an electric drive systemRequest a quote →
LMG671-ATE (1603.5886.02)Seven channels, no display, ATE version for an automated test lineRequest a quote →

Options / accessories

L60-CH-S3 (1603.5934.02) S channel expansion module: high accuracy, wide bandwidth, 0.015 % + 0.01 %, 18 bit, DC–10 MHz
L60-CH-A2 (1603.0132.02) A channel expansion module: high accuracy, wide bandwidth, 0.015 % + 0.01 %, 18 bit, DC–10 MHz
L60-CH-B3 (1603.5911.02) B channel expansion module: general purpose, 0.05 % + 0.02 %, 16 bit, DC–500 kHz
L60-CH-C2 (1603.5928.02) C channel expansion module: high accuracy at low frequency, 0.03 % + 0.01 %, 18 bit, DC–10 kHz
L60-CH-S-V (1603.1097.02) S channel voltage expansion adapter: up to 1500 V DC
L6-OPT-PSI (1603.0110.02) Process signal interface (PSI): 8 digital I/O, 2 high-speed analog inputs (150 kS/s), 8 low-speed analog inputs, 32 analog outputs, 2 frequency inputs
L6-OPT-CAN (1603.0849.02) CAN bus interface (ISO 11898-2)
L6X1-MS256 (1603.0910.02) 256 GB mass storage module (for measurement results)
L6X1-DVI (1603.0903.02) DVI interface (external monitor)
L6-OPT-SPV (1603.1200.02) Sampled value analysis option: 16 tracks streaming at 10 kS/s (up to 1.2 MS/s)
L6-OPT-SMV (1603.0890.02) Smart Vision: advanced convenience evaluation functions
L6-OPT-HRM (1603.0878.02) Harmonics analysis option: analysis up to the 2000th harmonic over the interface
L6-OPT-FLK (1603.0861.02) Flickermeter option (to EN 61000-4-15)
L6-OPT-EVT (1603.0855.02) Event trigger option: 16 tracks, 16 million samples per track
L6-OPT-SDC (1603.0826.02) Star-delta conversion option (needs 3 channels or more)
PCT200 (1603.1880.02) PCT200 precision current transformer: 200 Arms, 1 MHz bandwidth, ±0.0035 % accuracy
PCT600 (1603.1900.02) PCT600 precision current transformer: 600 Arms, 500 kHz bandwidth, ±0.0015 % accuracy
PCT1200 (1603.1874.02) PCT1200 precision current transformer: 1200 Arms, 400 kHz bandwidth, ±0.0015 % accuracy
PCT2000 (1603.1897.02) PCT2000 precision current transformer: 2000 Arms, 300 kHz bandwidth, ±0.0007 % accuracy
LT-CE-STBY (1603.1251.02) LMG Test Suite — standby power testing (to EN 50564/IEC 62301)
LT-CE-HRM (1603.1239.02) LMG Test Suite — harmonics testing (to EN 61000-3-2/-12)
LT-CE-FLK (1603.1222.02) LMG Test Suite — flicker testing (to EN 61000-3-3/-11)
PCT-DSUB2 (1603.1845.02) PCT sensor D-Sub cable (2 m)
PCTSIU4-1U (1603.1922.02) PCT sensor power supply (4 sensors, 19" 1 HU)
PCTX-K-L6 (1603.1951.02) PCT cable (2.5 m, for the LMG600)
L60-X-ADSE (1603.1100.02) LMG600 to PCT sensor adapter
L641-X-01 (1603.1122.02) LMG641 rack mount kit (19")
L671-X-01 (1603.1145.02) LMG671 rack mount kit (19")
L6X1-H2 (1603.1151.02) LMG671 front handles (2)
L6X1-H4 (1603.1168.02) LMG671 front and rear handles (4)

Frequently asked questions

Yes. KanTek Technology is the authorized Taiwan distributor for Rohde & Schwarz, supplying new R&S®LMG600 Series instruments with factory quotations and technical advice.

Instrument rental is launching soon. R&S®LMG600 Series is part of the line-up being prepared, and advance enquiries are welcome.

Yes. KanTek provides calibration and repair for R&S®LMG600 Series, with calibration performed by a partner laboratory, to keep the instrument accurate and available.

Use Request a quote on this page, or call +886-2-2999-7719. KanTek replies within 24 hours with pricing, lead time and the options that suit your application.

R&S®LMG600 Series is a Rohde & Schwarz product: High-precision power analyzer (DualPath architecture, 0.015 % of reading accuracy, 1/4/7 channels). Contact KanTek for selection or application advice.
Authorized R&S distributor
Repair and calibration services
Instrument rental, launching soon
Reply within 24 hours

Need R&S®LMG600 Series pricing, rental or calibration?

Request a quote